ARKRAY USA

ADAMS A1c HA-8180V

Superior Performance, Proven Technology

Improve patient care with the ADAMS™ A1c HA-8180V. The easy to use, automated hemoglobin A1c analyzer with diagnosis claims employs gold standard HPLC technology. The analyzer builds on proven technology to generate quality results, improve outcomes and provide superior precision to accurately report HbA1c.

Proven Technology

While new to the U.S. market, the ADAMS A1c brand has been used globally for decades. ARKRAY introduced the first fully automated HPLC hemoglobin analyzer in 1981 and has built upon the technology to meet the needs of today’s user.

Quality Results

The fully automated system utilizes gold standard HPLC technology. With CV’s of 1% or lower, the ADAMS A1c HA-8180V features the lowest NGSP precision claims on the market giving the laboratory confidence in generating accurate results for diagnosing and treating diabetes. The analyzer eliminates interference with the most common hemoglobin variants (HbS, HbC, HbD, and HbE) and features automatic sample mixing, assuring accurate and precise results.

Improve Outcomes

The ADAMS HA-8180V system allows physicians to accurately identify patients at risk, diagnose, and treat diabetes. Precise A1c results accurately monitor blood glucose control at critical decision making levels.

Easy to Use

The system is simple to operate and generates results in 90 seconds. Calibration and maintenance are minimized with the integrated column and pre-filter. Reduce hands-on time with primary tube cap-piercing sample and no sample pre-treatment required.

Product Brochure

Download Brochure

 
Fill out my online form.

Specifications

Measurement Items HbA1c (stable HbA1c, S-A1c)
HbF
Measurement Principle Ion-exchange High Performance Liquid Chromatography
Guaranteed Measurement Range HbA1c: 4 – 15% NGSP (20 – 140 mmol/mol)
HbF: 0.3 – 5%
Measurement Wavelength 420 nm / 500 nm (Dual-wavelength colorimetry)
Sample Type Whole blood or hemolysis
Sample Consumption Approximately 14 µL (whole blood)
Required Sample Volume Sample tube: 10 mm from bottom of tube
Sample cup: 400 µL
Analyzer Type Fully automated HbA1C analyzer
Throughput 37 tests/hour
Processing Speed 90 seconds/test
Sample Loading Capacity One-way transportation: 50 samples
Loop transportation: 100 samples
Sample Rack Capacity 10 samples/rack
Memory 900 results (including calibration results)
Communication RS-232 (Bi-directional)
Dimensions 21" (W) x 21" (D) x 21" (H)
Weight Analyzer: 86 lbs.
Sampler: 9 lbs.
Power Supply 100-240V, 50/60 Hz
Measurement Environment Temperature: 10 – 30° C
Humidity: 20 – 80% RH (no condensation)
Barcode specifications CODE 39, NW-7, ITF, CODE 128

SDS/MSDS

Coming soon

Product Brochure

Where to Buy

No matter the size of your laboratory, sample volumes processed, or supplier or group purchasing contracts, we have you covered.

Direct