The highly accurate GLUCOCARD® Shine family of blood glucose monitoring systems1 offer users both a full-size meter with a large LED screen and a compact system to compliment a user’s active lifestyle while retaining all the features they demand. The GLUCOCARD Shine XL has an ergonomic design developed to provide a secure grip during testing. The GLUCOCARD Shine’s small size is intended to improve portability whether the user is finishing up a workout or heading out on the town. Both systems offer a large screen intended to facilitate quick and easy viewing of blood glucose test results.

REFERENCE
1. ARKRAY. GLUCOCARD Shine XL Blood Glucose Monitoring System User Manual. 2017. 42-45.

Specifications

Test Strip Type GLUCOCARD Shine Test Strips
Test time 5 seconds
Meter range 20-600 mg/dL
Sample size 0.5 µL
Technology Plasma referenced
Calibration Plasma-equivalent
Display Back-lit LCD
Memory 500 tests with time & date stamp (Shine)
250 tests with time & date stamp (Shine XL)
Averaging 1-, 7-, 14-, 30- or 90-day averaging
Meter Size 3.74" x 1.3" x 0.75" (Shine)
3.66" x 1.85" x 0.59"
Weight 1.45 oz. (Shine)
1.82 oz. (Shine XL)
Power Battery: Two Single Replacement 3V (CR2032)
Tests/Battery 1,000 tests (Shine)
5,000 tests (Shine XL)

How-To

GLUCOCARD Shine XL User Overview Videos (English)

 
Meter Setup
 
Performing a Control Solution Test
 
Performing a Blood Glucose Test

GLUCOCARD Shine XL Videos de descripción general del usuario (Español)

 
Configuración del medidor
 
Realización de una prueba con solución de control
 
Realización de una prueba de glucosa en sangre

GLUCOCARD Shine XL Manuals and Guides / Manuales y Guías

User Manual
Manual de usuario
Quick Reference Guide / Guía de referencia rápida

GLUCOCARD Shine User Overview Videos / Videos de descripción general del usuario

 
User Overview Video
 
Video de descripción general del usuario

GLUCOCARD Shine Manuals and Guides / Manuales y Guías

User Manual
Manual de usuario
Quick Reference Guide / Guía de referencia rápida

Ordering Information

Name Product # UPC #015482- NDC #08317-
GLUCOCARD Shine Meter Kit 541100 54100-2 5411-00
GLUCOCARD Shine Meter Basic 1 540001 54001-2 5400-01
GLUCOCARD Shine XL Meter Kit 1 542110 54210-8 5421-10
GLUCOCARD Shine Test Strips 50 ct 540050 54100-2 5400-50

FAQ

Jump to GLUCOCARD® Shine FAQ
Jump to GLUCOCARD® Shine XL FAQ

GLUCOCARD® Shine FAQ

No. The BGMS and lancing device are intended for use by a single user only; never share the BGMS or lancing device with others.
No. Only GLUCOCARD® Shine Test Strips should be used with the GLUCOCARD Shine BGMS.
Yes. An abnormally high or low red blood cell count (hematocrit level over 60% or below 20%) may produce inaccurate results.
Two 3.0 V lithium batteries (disposable, type CR2032)
No. Reusing test strips will produce an inaccurate reading.
You should store your test strips in a cool and dry place at a temperature of 34-86ºF (1-30ºC). Keep your test strips away from direct sunlight or heat and do not freeze.
No. Store test strips in the product’s original vial. Close the vial tightly after taking out a test strip and use the test strip immediately.
No. Test strips used beyond the expiration date on the vial may produce inaccurate results.
Before setting the HYPo indicator, ask your doctor or diabetes nurse to help you decide what your hypoglycemia level will be.
Only test your meter and test strips using the GLUCOCARD® Shine Control Solutions (Control Level 1 and 2). These control solutions are specifically designed for this meter and test strips to ensure that they are working properly.
The control solution range is printed on your test strip vial. The ranges printed are for the GLUCOCARD Shine Control Solutions only.
Make sure your meter, test strips and control solution are at room temperature before testing. A control solution test must be done at room temperature 68-77ºF (20-25ºC).
Always make certain the control solution bottle is closed tightly and stored at a temperature of 46-86ºF (8-30ºC).
  • When you want to practice the test procedure using the control solution instead of blood
  • When using the meter for the first time
  • Whenever you open a new vial of test strips
  • If the meter or test strips do not function properly
  • If your symptoms are inconsistent with the blood glucose test results and you feel that the meter or test strips are not working properly
  • If you drop or damage the meter
  • When using the meter for blood glucose tests after any disinfection procedure
No. Repeat puncturing at the same sample site may cause pain or skin calluses (thick hard skin). Select a different site each time you test.
A “HI” reading appears when the blood glucose level is greater than 600 mg/dL and indicates severe hyperglycemia (much higher than normal glucose levels). If a “HI” reading occurs after retesting, contact your healthcare provider immediately. The meter otherwise displays results between 20-600 mg/dL.
A “Lo” reading appears when the result is less than 20 mg/dL and indicates severe hypoglycemia (very low glucose levels). If “Lo” is displayed again on retesting, please contact your healthcare professional immediately. The meter otherwise displays results between 20-600 mg/dL. When the hypoglycemia indicator is on, the symbol also appears.
500 glucose test results can be stored with time and date. Once the memory is full, the oldest test result will be deleted and the most current rest result will be stored.
It is recommended to disinfect your meter and lancing device at least once a week.
We have validated the Clorox Germicidal Wipes with 0.55% sodium hypochlorite as the active ingredient for disinfecting the meter and lancing device. Before disinfecting your meter and lancing device, read the instructions provided by the manufacturer of Clorox Germicidal Wipes. To purchase the disinfectant wipes, contact the manufacturer at www.cloroxprofessional.com or 800.537.1415.
Please review and follow the Pre-cleaning and Disinfection Procedures found in the user’s guide located with your Glucocard Shine BMGS kit. It is important that you use the validated wipes because other chemicals have not been validated and may damage the meter. Do not get fluids inside the meter through the test strip port, data transmission port or battery compartment. Never immerse the meter or hold it under running water because this will damage the meter.
The blood sample has abnormally high viscosity or insufficient volume. Repeat the test using a new test strip. For this error and a listing of other potential error codes, please refer to the GLUCOCARD® Shine User’s Manual.
Contact ARKRAY’s Customer Service Department (U.S. and Canada) at 800.566.8558.

GLUCOCARD® Shine XL FAQ

No. The BGMS and lancing device are intended for use by a single user only; never share the BGMS or lancing device with others.
No. Only GLUCOCARD® Shine Test Strips can be used with the GLUCOCARD Shine XL BGMS.
Yes. An abnormally high or low red blood cell count (hematocrit level over 60% or below 20%) may produce inaccurate results.
Two 3.0 V lithium batteries (disposable, type CR2032)
No. Reusing test strips will produce an inaccurate reading.
You should store your test strips in a cool and dry place at a temperature of 34-86ºF (1-30ºC). Keep your test strips away from direct sunlight or heat, and do not freeze.
No. Store test strips in the product’s original vial. Close the vial tightly after taking out a test strip and use the test strip immediately.
No. Test strips used beyond the expiration date on the vial may produce inaccurate results.
Only test your meter and test strips using the GLUCOCARD® Shine Control Solutions (Level 1 and 2). These control solutions are specifically designed for this meter and test strips to ensure that they are working properly.
The control solution range is printed on your test strip vial. The ranges printed are for the GLUCOCARD Shine Control Solutions only.
Make sure your meter, test strips and control solution are at room temperature before testing. A control solution test must be done at room temperature 68-77ºF (20-25ºC).
Always make certain the control solution bottle is closed tightly and stored at a temperature of 46-86ºF (8-30ºC).
  • When you want to practice the test procedure using the control solution instead of blood
  • When using the meter for the first time
  • Whenever you open a new vial of test strips
  • If the meter or test strips do not function properly
  • If your symptoms are inconsistent with the blood glucose test results and you feel that the meter or test strips are not working properly
  • If you drop or damage the meter
  • When using the meter for blood glucose tests after any disinfection procedure
No. Select a different site each time you test. Repeat puncturing at the same sample site may cause pain or skin calluses (thick, hard skin).
A “HI” reading appears when the blood glucose level is greater than 600 mg/dL and indicates severe hyperglycemia (much higher than normal glucose levels). If a “HI” reading occurs after retesting, contact your healthcare provider immediately. The meter otherwise displays results between 20-600 mg/dL.
A “Lo” reading appears when the result is less than 20 mg/dL and indicates severe hypoglycemia (very low glucose levels). If “Lo” is displayed again on retesting, contact your healthcare professional immediately. The meter otherwise displays results between 20-600 mg/dL.
250 blood glucose test results can be stored with time and date. Once the memory is full, the oldest test result will be deleted and the most current test result will be stored.
Use a soft cloth or tissue to wipe the meter exterior. If necessary, dip the soft cloth or tissue in a small amount of alcohol. Do NOT use organic solvents such as benzene, acetone, or any household and industrial cleaners that may cause irreparable damage to the meter.
Please review and follow “Caring for the GLUCOCARD Shine XL System” found in the user manual located with your GLUCOCARD Shine BGMS kit. Do not get fluids inside the meter through the test strip port, data port or battery compartment. Never immerse the meter or hold it under running water because this will damage the meter.
The blood sample has abnormally high viscosity or insufficient volume. Repeat the test using a new test strip. For this error and a listing of other potential error codes, please refer to the GLUCOCARD Shine XL User Manual.
Contact Customer Service at 800.566.8558, 24 hours a day, 7 days a week, U.S.A. and Canada.